Vibration isolation solution for Zeiss sigma500 electron microscope

2025-08-27 14:44:28 admin

Background

User: Harbin Institute of Technology

This research laboratory mainly focuses on water treatment and the field of water films

Instrument: ZEISS Sigma-500

Field emission scanning electron microscopes for quality imaging and advanced analysis

Vibration isolation solution supplier: Unicon Technology Co., LTD

+ Environmental solution provider

+ Research institution service provider

+ Table Stable Active vibration isolation system supplier

Description

The external environment of the laboratory is rather complex. Multiple research rooms are placed side by side on the same floor, with a high density of people and goods coming and going. Within a 10-meter range outside, the pump room and the distance from the elevator shaft to the SEM tube are less than 5 meters. As the horizontal level of the elevator shaft is lower than the laboratory floor and it is a trough frame structure, when the two boxes of the elevator work freely up and down, they will generate significant vibrations, which directly act on the tube of the electron microscope and affect its ability to capture images normally.

The Sigma500 integrates an air cushion inside, providing excellent shock absorption within the frequency range of 25Hz to 40Hz. The main purpose is to slow down the high-frequency mechanical vibrations generated inside the electron microscope, but it cannot eliminate the low-frequency vibrations from the ground.

The active vibration isolation system can perform active shock absorption within the range of 0.7Hz to 200Hz thereby providing a stable environment for the electron microscope and enhancing the quality of the captured images.

Picture Keywords

Picture Keywords

Table Stable active vibration isolation technology
Both active and passive isolation mechanisms work simultaneously to produce vibration isolation effects
Piezoelectric ceramic technology detects incoming vibrations and actively eliminates them
+ No low-frequency resonance

It has excellent wide-spectrum attenuation capability in all six degrees of freedom in space

"Rigid design

AVI-400 Active vibration isolation system
"Modular design"
+ Has load capacity
+ Small size design
+ Applicable to electron microscopes of all brands


Step 1: Lift the electron microscope as a whole

Step 2: Place the vibration isolation system

Step 3: Connect the lines and adjust the load

Step 4: Start using

Picture Keywords

Installation time: Approximately 1 hour

Required tools: Special lifting system for electron microscopes & Adjustable wrench & Hex socket wrench &" Special open-end wrench

Picture Keywords

Data display

Picture Keywords

Image display

300,000 time-image obtained when the elevator is in operation vs. 300,000 time-image obtained when the elevator is in operation and actively isolated

Picture KeywordsPicture Keywords


The image obtained after active vibration isolation

Picture Keywords

Advantage

"Matching electron microscope"

Miniaturization, suitable for personalized requirements

The module is expandable

+ Does not increase the horizontal space of the electron microscope

The thin design increases the center of gravity of the electron microscope by only 110mm

It is easy to install and debug, and no foreign engineers are required for installation

+ Low maintenance costs

+ Domestic prototype trial and demonstration